  • Internal/Mapping/Calibration

    v14 v15  
    1616 1. Pick up one node from the grid and take it down from the shelf
    1717 1. load baseline-calibration.ndz image to the node
    18  1. Connecting RF cable to one of the card
     18 1. Connecting ESG RF cable to one of the card
    1919 1. ssh to console.grid
    20  1. ./caliber.pl [mac-addr]
    21  1. check the calibration_central database in idb1.
     20 1. Run Card-rx experiment '''./cardcalibrate.pl 21 1 [cardid] [mac-addr]'''
     21 1. disconnect the ESG outlet from the card. Connect the card to VSA RF outlet
     22 1. Run card-tx experiment '''./cardtxcalib.pl [mac-addr]'''
     23 1. check the '''calibration_central''' database in ''idb1''.
    2224 1. Do above procedures for another card